A force probe inside an electron microscope
Byeonghoon Lim builds quartz tuning fork AFM probes at Jeonbuk National University.
| Date | Meeting | Form | Subject |
|---|---|---|---|
| 2026.10 | IEDMS 2026, Taipei | Oral | Drift correction |
| 2026 | ISPSA 2026 | Poster | Drift correction |
| 2025.12 | 1st Basic Science Festival | Poster | Three surfaces |
| 2025.11 | Korean Society of Surface Analysis | Poster | Three surfaces |
| 2025.08 | IWSPM 2025 | Poster | Charged surfaces |
Quartz fork, gold-coated tip
A 32,768 Hz fork, read electrically, so no laser enters the chamber.
Vacuum doubles the Q-factor
A tipped fork swept through resonance in air, then in the evacuated chamber.
- Vacuum, Q-factor 20,319
- Air, Q-factor 9,568
Amplitude, V
Detuning from the peak, Hz
| Environment | Q-factor | Peak, Hz | Amplitude, V | Drive, mV rms |
|---|---|---|---|---|
| Air | 9,568 | 32,534.2 | 6.912 | 130 |
| Vacuum | 20,319 | 32,542.2 | 6.714 | 60 |
How the tip is made
Pull
A quartz rod is heated with a laser and pulled to a point.
Coat
The point is coated with gold.
Mount
The rod is attached to one prong of the fork.
Drift between approach and retraction
On silicon in vacuum, the two curves disagree in position at equal amplitude.
- −3.068 nm
- Mismatch at 4.0 V in the raw curves
- −0.350 nm
- The same mismatch after alignment
- +3.098 nm
- Offset applied to every retraction point
Presented at IEDMS 2026. The residual still varies with amplitude level.
Three surfaces under the beam
Mica
A dielectric. Damping appears with the beam on.
Silicon
A semiconductor. Curves in vacuum and under the beam nearly coincide.
Gold
A conductor. Damping appears under the beam, as on mica.
Manuscript comparing the three in preparation.
Where to find me
- Lab
- Microscope-based Nanoscale Physics Lab, Department of Physics, Jeonbuk National University, Jeonju, Korea
- Advisor
- Prof. Sangmin An
- Next
- A master’s in physics from 2027, toward silicon MEMS processing
- limbh8218@gmail.com